Agilent Technologies 8510C Manuel de service Page 69

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 222
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 68
8510XF Network Analyzer Systems 2-41
Installation
System Operational Test
Conclusion If all of the measured levels are at or above the minimum levels shown in
Figure 2-14, this indicates that the 8510XF is functioning normally, and is
ready to be used.
NOTE: frequencies below 2 GHz are rolled off sharply, owing to the characteristics of the directional
couplers in the test heads.
Table 2-14 Minimum Levels (All User Parameters)
Frequency Range (GHz) Minimum Level (dBm)
<2 63
2 to <18 38
18 to <50 58
50 to <75 61
>75 63
Vue de la page 68
1 2 ... 64 65 66 67 68 69 70 71 72 73 74 ... 221 222

Commentaires sur ces manuels

Pas de commentaire