Vue de la page 93
3-24 8510XF Network Analyzer Systems
Operation
Standard Types
One-Path
Transmission
Standards
These standards are used to measure transmission and match for one-path
2-port calibrations. When you perform a calibration of that type using a
1.0 mm calibration kit, you will see the following menu:
Figure 3-12 One-Path Transmission Calibration Menu
& Equivalent GPIB Commands
Notes on the Standards {FWD.TRANS.
THRU}
Make a “thru” connection (that is, connect one port to the
other through a cable).
{FWD.MATCH
THRU}
Make a “thru” connection (that is, connect one port to the
other through a cable).
NOTE It is necessary to make both of these measurements.