Relay Multiplexer Command Reference 53
DISPlay
The DISPlay subsystem monitors the channel state of a selected module (or card) in
a switchbox, and only operates with mainframes that have a display, such as the
75000 Series B Mainframe (Model Number E1301).
Subsystem Syntax DISPlay
:MONitor
:CARD <
number|AUTO>
[:STATe] <
mode>
:MONitor:CARD
DISPlay:MONitor:CARD <number|AUTO> selects the module in a switchbox to be
monitored.
Parameters
Comments Selecting a Specific Module to be Monitored: Send the card number in a
switchbox with the
DISPlay:MONitor:CARD command.
Selecting the Present Module to be Monitored: Use the
DISPlay:MONitor AUTO command to select the last module addressed
by a switching command (e.g.,
[ROUTe:]CLOSe).
*RST Condition: DISPlay:MONitor:CARD AUTO
Example Select Module #2 in a Switchbox for Monitoring
DISP:MON:CARD 2
Selects module #2 in a switchbox.
Parameter Name Parameter Type Range of Values
<
number|AUTO>
numeric 1–99