56 Relay Multiplexer Command Reference
:CONTinuous?
INITiate:CONTinuous? queries the scanning state. With continuous scanning
enabled, the command returns
ON. With continuous scanning disabled, the
command returns
OFF.
Example Query Continuous Scanning State
INIT:CONT ON
Enables continuous scanning.
INIT:CONT?
Query continuous scanning state.
[:IMMediate]
INITiate[:IMMediate] starts the scanning cycle and closes the first channel in the
channel list. Successive triggers from the source specified by the
TRIGger:SOURce
command advances the scan through the channel list.
Comments Starting the Scanning Cycle: The INITiate[:IMMediate] command starts
scanning by closing the first channel in the channel list. A trigger advances the
scan through the channel list. An invalid channel list generates an error
(see the
[ROUTe:]SCAN command).
Stopping Scanning Cycles: See the ABORt command.
Example Enabling a Single Scan
SCAN (@100:115)
Sets channel list.
INIT
Starts scanning cycle.