Calibrating the Infiniium Oscilloscope and Probe 14
U7232A DisplayPort Compliance Testing Notes 145
c Follow the on- screen instructions.
d At the end of the Atten/Offset Calibration perform the Skew
Calibration.
SMA Probe Head Skew Calibration
This procedure ensures that the timing skew errors between channels are
minimized. Perform the following steps:
1 Referring to Figure 63 below, perform the following steps:
a Select the Start Skew Calibration button and follow the on- screen
instructions. For more information on proper connection of probe to
the oscilloscope, refer to the De- skew and Calibration manual. This
Figure 62 Probe Calibration Window.