DisplayPort Test Implementation Group 24 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Figure 21: J-BERT N4903B Differential Amplitude Adjustment
b. In the oscilloscope, using the Analysis > Serial Data > Clock Recovery menu, change the
Clock Recovery from Constant Frequency to 2
nd
Order PLL.
Figure 22: DSA 2
nd
Order PLL Clock Recovery
c. On the Oscilloscope, Select Measure > Eye > Eye Height Measurement.
d. Alternate between adjusting the Vampl on the J-BERT N4903B user interface and pressing
the Single Acquisition Button on the Oscilloscope until the Eye Height is calibrated to the
following values.
For RBR
For HBR
For HBR2
Adjust Eye Height to
46mV
diff
+/-10%
Repeat measurement at least 5
times and average the results.
See C appendix
Adjust Eye Height to 150mV
diff
+/-10%
Repeat measurement at least 5
times and average the results.
Adjust Eye Height to 120mV
diffI
+/-10% as shown in the below
example
Repeat measurement at least 5
times and average the results.