DisplayPort Test Implementation Group 49 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Automated Sink Calibration
Calibration steps are grouped by data rates. RBR and HBR calibration consist of
ISI verification
RJ calibration
Eye height calibration
HBR2 calibration of
ISI verification
RJ calibration
Fixed SJ at 200MHz calibration
Eye height calibration without crosstalk
Eye height calibration with cross talk applied
All calibration steps are CTS compliant.
Figure 49: Calibration step selection
Several calibration steps can be selected by check marks and the test automation will execute the selected
calibration steps sequentially. The user gets prompted only in the beginning and when a setup change is
necessary.