DisplayPort Test Implementation Group 44 Agilent J-BERT N4903B & 90000A/X/Q Series Sink Draft2
Figure 41: RBR ISI measurement step 2 for jitter with ISI
In this example the hist
pp no ISI
measurement was 58.9ps and the hist
pp with ISI
measurement was 406.7ps.
The delta is 347.8ps or 563.4mUI. The target limits are 570mUI+-5%.
Eye Height Measurement for RBR
A vertical histogram which slices through the one level and one for the zero level are used to measure eye
height boundaries for the zero level and one level. The delta is the eye height. The scope is triggered by a
signal from J-BERT’s clock or trigger output.
Scope settings:
• Set scope up for a measurement on a differential signal either by using a differential SMA probe head
or HW based channel subtraction (channel 1-3 or channel 2-4)
• Use 1kpts acquisition depths
• Set display mode to color grade
• Adjust eye in the center of the display
• Activate a vertical histogram Analyze ! Histogram and slice thinly through the one level starting in
the center of the eye for the one level boundary and change the window in the voltage axis to slice
through the zero level starting from the center of the eye for the zero level boundary
• Activate an histogram peak to peak measurement
The measurements for one level and zero level boundaries need to be done with the same left and right
limits for the histogram window to get the eye height for a specific timing setting.