24 Getting Started Guide
1 Quick-Start
Port Configuration
Any of the TRXs four RF ports can be configured as an input port (accepting a signal to
be analyzed) or as an output port (providing an RF stimulus). Two of the ports (RFIO 1
and RFIO 2) can also be used in full-duplex mode (that is, with a single port used for
both input and output simultaneously). The various possible configurations are described
here.
Full Duplex Port Setup (RFIO 1 and RFIO 2)
Ports RFIO 1 and RFIO 2 connect to a splitter/combiner rather than a switch, so they can
be used in full duplex mode.
To use the RFIO 1 port as the input and output, set up the RF Input using this menu
setting or its equivalent command:
Input/Output > RF Input > RF Input Port > RFIO 1
Command:
:FEED:RF:PORT RFIO1
Then set up the RF Output using this menu setting or its equivalent command:
Input/Output > More > RF Output & Test Set Config > RF Output Port > RFIO 1
Command:
:FEED:RF:PORT:OUTP RFIO1