CONFigure:VOLTage:DC {<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for dc voltage measurements with
the specified range and resolution. This command does not initiate the
measurement.
CONFigure:VOLTage:DC:RATio {<
range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for dc:dc ratio measurements with
the specified range and resolution. This command does not initiate the
measurement. For ratio measurements, the specified range applies to
the signal connected to the
Input terminals. Autoranging is automatically
selected for reference voltage measurements on the
Sense terminals.
CONFigure:VOLTage:AC {<
range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for ac voltage measurements with
the specified range and resolution. This command does not initiate the
measurement. For ac measurements, resolution is actually fixed at
6
1
2
digits. The resolution parameter only affects the front-panel display.
CONFigure:CURRent:DC {<
range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for dc current measurements with
the specified range and resolution. This command does not initiate the
measurement.
CONFigure:CURRent:AC {<
range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for ac current measurements with
the specified range and resolution. This command does not initiate the
measurement. For ac measurements, resolution is actually fixed at
6
1
2
digits. The resolution parameter only affects the front-panel display.
CONFigure:RESistance {<
range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for 2-wire ohms measurements
with the specified range and resolution. This command does not initiate
the measurement.
CONFigure:FRESistance {<
range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
Preset and configure the multimeter for 4-wire ohms measurements
with the specified range and resolution. This command does not initiate
the measurement.
4
Chapter 4 Remote Interface Reference
The MEASure? and CONFigure Commands
119