Module 7
Measurement Functions
7-42
Calculated value
1.5 M ohm (with Series Resistor)
SMU Series Resistor
Compensation
If you use the SMU series resistor, you will need to compensate the measurement data to eliminate
the effects of the series resistor (1 M ohm).
This is a compensation example by using the following user function.
Vcalc = V1 – I1 * 1000000
The Rcalc value is obtained by using the analysis function and the auto analysis function as same as
the IV-res setup.
This compensation example is shown in the Rcalc setup.