2-4 Agilent 4155B/4156B Quick Start Guide, Edition 2
Making a Measurement
Step 1. Prepare for the measurement
Step 1. Prepare for the measurement
Before executing measurement, configure the 4155B/4156B and accessories.
1. Make sure that the 4155B/4156B is off.
2. Connect the 16442A test fixture to the 4155B/4156B. See next figure.
3. If you use the keyboard, connect it to the 4155B/4156B.
4. If you use the 4156B, connect as shown below:
4156B cable 16442A 4156B cable 16442A
Intlk
Interlock/LED
a
a. Interlock/LED cable: Agilent 16493J
Intlk VSU 1
Coaxial
b
b. Coaxial cable: Agilent 16493B. You do not need to connect VSUs and
VMUs for this measurement.
VSU 1
SMU 1
Kelvin triaxial
c
c. Kelvin triaxial cable: Agilent 16493K
SMU 1 VSU 2
Coaxial
b
VSU 2
SMU 2
Kelvin triaxial
c
SMU 2 VMU 1
Coaxial
b
VMU 1
SMU 3
Kelvin triaxial
c
SMU 3 VMU 2
Coaxial
b
VMU 2
4156B
16442A
Kelvin Triaxial Cables
Coaxial Cables
Interlock/LED Cable