1-116
Making Measurements
Using Test Sequencing to Test a Device
This will create the following displayed lists:
Start of Sequence
LOOP COUNTER
7x1
INTERNAL DISK
DATA ONLY
ON
DO SEQUENCE
SEQUENCE 2
Start of Sequence
FILE NAME
DT[LOOP]
PLOT NAME
PL[LOOP]
SINGLE
SAVE FILE 0
PLOT
DECR LOOP COUNTER
IF LOOP COUNTER <> 0 THEN DO
SEQUENCE 2
Sequence 1 initializes the loop counter and calls sequence 2. Sequence 2 repeats until the
loop counter reaches 0. For each loop, it takes a single sweep, saves the data file and plots
the display.
Save/Recall
FILE UTILITIES
SEQUENCE FILE NAMING
FILE NAME FILE0
ERASE TITLE
D
T
LOOP COUNTER
DONE
PLOT NAME PLOTFILE
ERASE TITLE
P
L
LOOP COUNTER
DONE
RETURN
Sweep Setup
TRIGGER MENU
SINGLE
Save/Recall
SAVE STATE
Copy
PLOT
Seq
SPECIAL FUNCTIONS
DECISION MAKING
DECR LOOP COUNTER
IF LOOP COUNTER 0
SEQUENCE 2 SEQ 2
Seq
DONE SEQ MODIFY
Vue de la page 128