Making Mixer Measurements (Option 089 Only)
Fixed IF Mixer Measurements
Putting the Analyzer into Tuned Receiver Mode
Setting Up a Frequency List Sweep of 26 Points
Performing a Response Calibration
Input as title: POW:LEV 6DBM
Input as title: FREQ:MODE CW;CW 100MHZ
Prompting the User to Connect a Mixer to the Test Setup
Input as title: CONNECT MIXER
Local
SYSTEM CONTROLLER
System
INSTRUMENT MODE
TUNED RECEIVER
Sweep Setup
SWEEP TYPE MENU
EDIT LIST
ADD
START
100 M/µ
STOP
100 M/µ
NUMBER OF POINTS
26 x1
DONE
DONE
LIST FREQ
Meas
INPUT PORTS B
Display
MORE
TITLE
ERASE TITLE
DONE
Seq
SPECIAL FUNCTIONS
PERIPHERAL HPIB ADDR
HIB ADDR
19 x1
TITLE TO PERIPHERAL
Display
MORE
TITLE
ERASE TITLE
DONE
Seq
SPECIAL FUNCTIONS
PERIPHERAL HPIB ADDR
HIB ADDR
19 x1
TITLE TO PERIPHERAL
Cal
CALIBRATE MENU
RESPONSE
THRU
Display
MORE
TITLE
ERASE TITLE
DONE
Seq
SPECIAL FUNCTIONS
PAUSE