Chapter 7 451
Language Reference
DEMODT Demodulation Time
DEMODT Demodulation Time
Syntax
Figure 7-91 DEMODT Syntax
Description
The DEMODT command selects the amount of time that the sweep
pauses at the marker to demodulate a signal. The default value is 1
second. When the frequency span equals 0 Hz, demodulation is
continuous, except when between sweeps. For truly continuous
demodulation, set the frequency span to 0 Hz and the trigger mode to
single sweep (see TM).
Parameters
number real from 100E3 to 60.
UP or DN increments in a 1, 2, 5, 10 sequence.
Preset State
Off