638 Chapter7
Language Reference
TRA/TRB Trace Data Input/Output
TRA/TRB Trace Data Input/Output
Syntax
Figure 7-281 TRA/TRB Syntax
Description
The TRA and TRB commands provide a method for transferring trace
data to or from a computer. The available data formats are parameter
(P) format, binary (B) format, A-block format, I-block format, or
measurement units (M) format. Transfers to the computer must be
completed within 30 seconds or the transfer will be aborted. For
instructions on transferring data and more examples, refer to Chapter
5, "Programming."
Option 007, Fast Digitized Time Domain, is necessary to obtain trace
data with sweep times <30 ms. Without Option 007, sweep times <30
ms are not digitized. If a sweep time <30 ms is selected and Option 007
is not installed, the trace data returned is the last trace data taken with
a sweep time 30 ms.
Parameter
msb and lsb length represents the length of the trace as two 8-bit bytes.