Cursor Measurement Hints
If cursors are positioned too closely together, an error will be displayed when
the SET softkey is selected.
Displayed marker readings in percent (%) and degrees (°) are always relative
measurements, with the current reading dependent on the previously
established 100% or 360 reference setting.
Time Cursor Measurements in Degrees
Figure 2 11
Operating the Measurement/Storage Module
To make additional cursor measurements
2–28