Agilent Technologies EasyEXPERT Guide de l'utilisateur Page 318

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 344
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 317
4- 74 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Function Details
Initial Settings
Initial Settings
Agilent B1500 is initialized by turning it on, the *RST command, or the device
clear. Initial settings of the B1500 are shown in the following tables.
Table 4-8 Initial Settings
Setup Item Initial Setting
Mainframe Auto calibration off
SMU Output switch close
Filter on
Series resistor, only for HR/MP/HP off
ASU path/indicator SMU side/enable
ASU 1 pA auto range disable
Current measurement range with pulse compliance range
without pulse auto
Voltage measurement range with pulse compliance range
without pulse auto
A/D converter HR/MP/HPSMU: high resolution ADC
HC/DHC/MC/HVSMU: high speed ADC
ADC integration time high speed ADC: auto
high resolution ADC: auto
ADC zero function off
Sweep automatic abort function off
Hold time 0 s
Delay time 0 s
Pulse width 0.001 s
Pulse period 0.01 s
Vue de la page 317
1 2 ... 313 314 315 316 317 318 319 320 321 322 323 ... 343 344

Commentaires sur ces manuels

Pas de commentaire