1- 42 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Main GUI
Tracer Test
Oscilloscope View
This window is opened/closed by clicking the Oscilloscope View button on the
Tracer Test screen, or selecting the Option > Oscilloscope View menu.
This function monitors a point of the pulsed sweep measurement or the pulsed spot
measurement and displays its waveform. The monitoring point is specified by a
rectangular marker on the trace. This function is used for monitoring the time
relationship between multiple channel outputs/measurements, the change of the
output level for the gate drive channel, and so on.
This function is valid for using one or more VPULSE or IPULSE channels. And this
function is not available for the following measurement resources.
HRSMU, HPSMU, MPSMU
•MFCMU
•WGFMU
SPGU
GNDU