Agilent Technologies EasyEXPERT Guide de l'utilisateur Page 70

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 344
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 69
1- 58 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Main GUI
Calibration
Calibration
The Calibration window is opened by clicking the Calibration button, and is used to
perform the self-calibration, SMU current offset cancel, CMU data correction, and
so on.
“Module Self Calibration”
“SMU Zero Cancel”
“CMU Calibration”
Module Self Calibration
The Module Self Calibration tab screen is used to perform the self-calibration of
modules. Check the left check box of Name to select the module for self-calibration.
Open the measurement terminals at the cable end before performing the
self-calibration.
Module Self
Calibration
Displays the following values.
Name Name of module.
Full Range Full range calibration On or Off. Only for SMU+ASU.
Status Calibration status. Pass or fail.
Notes Error information when calibration failed.
Select All Selects all modules for calibration. This checks the check box for all modules.
Unselect All Unselects all modules. This unchecks the check box for all modules.
Start Calibration Starts calibration for the selected modules.
Abort Aborts calibration.
Full Range
Calibration
Only for the B1500A. Check this box to enable the 1 pA range calibration of
SMU+ASU.
If you do not use the 1 pA range, uncheck this box. The B1500A does not perform
the 1 pA range offset data measurement and calibration.
Vue de la page 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 343 344

Commentaires sur ces manuels

Pas de commentaire