SSB Receivers
180
On the Test Set using the knob and data entry keys:
16. Change the value of RF Gen Freq until the AF Freq meter reads 1 kHz.
On the Radio set the Receiver’s Controls as follows:
17. Set the Receiver’s volume control until the AC Level meter indicates the
Receiver’s rated output power.
Refer to audio output specifications for the Receiver being tested as
required. If unknown, increase the volume control until the AC Level
meter stops increasing (or at maximum).
On the Test Set using the knob and data entry keys:
18. Set AF Freq meter to measure SINAD.
19. Select AC Level.
20. Press the REF SET key.