Agilent Technologies HP8648A Manuel d'utilisateur Page 270

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 320
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 269
AF Analyzer Specifications
269
Distortion Measurement
Fundamental Frequency: 1 kHz ±5 Hz
Optional Frequency Range: 300 Hz to 10 kHz ±5% (Option 019)
Input Level Range: 30 mV to 30 Vrms
Display Range: 0.1% to 100%
Accuracy:
±1 dB (0.5 to 100% distortion) for tones from 300 to 1500 Hz measured
with the 15 kHz LPF
±1.5 dB (1.5 to 100% distortion) for tones from 300 Hz to 10 kHz mea-
sured with the >99 kHz LPF
Residual THD+Noise:
60 dBc or 150 µV, whichever is greater, for tones from 300 to 1500
Hz measured with the 15 kHz LPF
57 dBc or 450 µV, whichever is greater, for tones from 300 Hz to 10
kHz measured with the >99 kHz LPF
Resolution: 0.1% Distortion
SINAD Measurement
Fundamental Frequency: 1 kHz ±5 Hz
Optional Frequency Range: 300 Hz to 10 kHz ±5% (Option 019)
Input Level Range: 30 mV to 30 Vrms
Display Range: 0 to 60 dB
Accuracy:
±1 dB (0 to 46 dB SINAD) for tones from 300 to 1500 Hz measured
with the 15 kHz LPF
±1.5 dB (0 to 36 dB SINAD) for tones from 300 Hz to 10 kHz measured
with the >99 kHz LPF
Residual THD+Noise:
60 dB or 150 µV, whichever is greater, for tones from 300 to 1500 Hz
measured with the 15 kHz LPF
57 dBc or 450 µV, whichever is greater, for tones from 300 Hz to 10
kHz measured with the >99 kHz LPF
Resolution: 0.01 dB
Vue de la page 269
1 2 ... 265 266 267 268 269 270 271 272 273 274 275 ... 319 320

Commentaires sur ces manuels

Pas de commentaire