FM Transmitters
68
Measurement Procedure:
1 Connect the Transmitter as shown.
CAUTION: The RF present at the Test Set RF IN/OUT connector must not exceed 60W
continuous (or 100 Watts for 10 sec/minute).
On the Test Set:
2 Press the PRESET key.
Using the knob and the data entry keys:
3 Select the AF ANL screen.
4 Set Filter 1 to <20 Hz HPF setting.
5 Set Filter 2 to 300 Hz LPF setting.
6 Set Settling to Slow setting.
7 Select the DECODER screen.
8 Set Mode to CDCSS.
9 Set Standard to CDCSS.
10 Set Input Level to 0.95 kHz.
11 Set Arm Meas to Cont.
On the Radio:
12 Key the Transmitter and keep keyed until the remaining steps are complet-
ed.
Data rate, binary data (bin), and the Octal Code(s) are displayed on the
Test Set as shown.
Test Set Options Required
Decoder Option