4-10 Agilent B2200 Users Guide, Edition 2
Programming
Programming Examples
Using Bias Mode
The following example uses the bias mode.
Figure 4-3 Bias Mode Example
Setup: Channel configuration mode: Auto
Connection rule: Single. Multiple connection is available for the bias port.
Connection sequence: Break_Before_Make
Used module: All switch modules installed in the mainframe.
Bias port: AUX Input 10
Bias enabled output port: All output ports
Connection paths (connected when the bias mode is ON):
from bias port to all bias enabled output ports
Connection paths (connected by the :ROUT:CLOS command):
The following output ports will be disconnected from the bias port, and the
following input-output connections will be made.
from SMU1 to Output 1 (channel list 00101)
from SMU2 to Output 2 (channel list 00202)
from SMU3 to Output 3 (channel list 00303)
from SMU4 to Output 4 (channel list 00404)