Agilent Technologies Agilent B2201A Manuel d'utilisateur Page 137

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 234
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 136
Agilent B2200 Users Guide, Edition 2 5-11
SCPI Command Reference
*STB?
The following is example for query:
OUTPUT @Agb2200;"*SRE?"
ENTER @Agb2200;A
*STB?
This query command reads the Status Byte Register (reads Master Summary Status
bit, not Request for Service Message).
For bit6, this command reads MSS, not Request for Service (RQS). See “Status
Reporting Structure” on page 5-53.
Syntax *STB?
Query response register <newline><^END>
Semantics The following table shows the bits of the Status Byte Register, and the
binary-weighted decimal value of each bit.
Example OUTPUT @Agb2200;"*STB?"
ENTER @Agb2200;A
Parameter Explanation
register decimal integer value that is the sum of the binary-weighted values
for the set bits
bit binary-weight description
0 1 not used. always 0.
1 2 not used. always 0.
2 4 not used. always 0.
3 8 not used. always 0.
4 16 MAV (Message Available summary-message)
5 32 ESB (Event Status Bit)
6 64 MSS (Master Summary Status)
7 128 not used. always 0.
Vue de la page 136
1 2 ... 132 133 134 135 136 137 138 139 140 141 142 ... 233 234

Commentaires sur ces manuels

Pas de commentaire