Agilent Technologies 1100 Series Manuel d'instructions Page 169

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 196
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 168
1100 Series Capillary LC System Manual 159
Options 6
9 Insert the PEEK capillaries from the flow cell body
together with the SST fitting, ferrule and cone into
the union and tighten it.
Remove the flow cell and perform a leak test.
If no leak is observed, install the flow cell and you are
ready to work.
Make sure that the flow cell assembly is inserted
correctly and fits perfectly in the optical unit
(especially when PEEK capillaries are used).
Vue de la page 168
1 2 ... 164 165 166 167 168 169 170 171 172 173 174 ... 195 196

Commentaires sur ces manuels

Pas de commentaire