Agilent Technologies 1100 Series Manuel d'instructions Page 173

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 196
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 172
1100 Series Capillary LC System Manual 163
Options 6
9 Insert the flow cell capillaries and tighten them
fingertight.Use of the wrench and torque adapter as
described page 155 and tighten the fittings alternate
10 Reassemble the flow cell body to the holder.
11 Re-install the flow cell and connect the capillaries to
the union holder.
Remove the flow cell and perform a leak test.
If no leak is observed, install the flow cell and you are
ready to work.
IN
OUT
Vue de la page 172
1 2 ... 168 169 170 171 172 173 174 175 176 177 178 ... 195 196

Commentaires sur ces manuels

Pas de commentaire