162 1100 Series Capillary LC System Manual
6Options
5 The quartz body and the cell seal assembly can be
separated for cleaning purpose.
6 This figure shows the correct holding of the quartz
body and the cell seal assembly.
7 Replace the cell seal assembly onto the quartz body.
It is recommended to use a new assembly.
8 Slide the quartz body completely into the cell body to
the front stop (use for example a toothpick).