54 1100 Series Capillary LC System Manual
3 Capillaries and Fittings
Hints for Successful use of Capillaries and Fittings
Never overtighten a fitting.
Never cut a capillary
Take care when you bend a capillary (diameter never below 40 mm).
Avoid the use of alkaline solutions (pH > 8.5) which can attack the fused
silica from the capillaries.
When connecting, press the capillary smoothly into the fitting to avoid air
gaps.
If a capillary leaks, never retighten it under flow.
A blocked capillary can often be cleaned by flushing it back. Acetone is
recommended for this.
Take care when applying module doors, not to crush capillaries.
A broken capillary can release silica particles in the system.