Chapter 3 47
Configuration
Configuration Methods
10 At the Telnet prompt (>), enter the configuration values you want to
change or set. (You only need to specify the non-default values that
you want.) The following information will help you set the values
correctly.
The following table lists the configuration values you can set using
Telnet. The following information shows the syntax you should use
and provides tips to help you to change or set the configuration
values correctly.
NOTE
By using the Telnet configuration method, the LAN hardware address
(hardware-addr:) value is automatically set to the correct hardware
address for this LAN/GPIB Gateway. Do not change the hardware
address value via Telnet. Its value in the Telnet configuration table is for
information purposes only.
Configuration Value Telnet Command Default Value
Hostname hostname: E2050
IP Address ip: 192.0.0.192
Default Subnet Gateway Address gateway: 0.0.0.0
Subnet Mask subnet-mask: 0.0.0.0
Syslog Server Address syslog-svr: 0.0.0.0
BOOTP bootp: ON
LAN Timeout lan-timeout: 0
I/O Timeout io-timeout: 120
IP Allow List allow: * (All allowed)
GPIB Address hpib-address: 21
GPIB Interface Name hpib-name:
hpib
1
GPIB Logical Unit hpib-unit: 7
1