Agilent Technologies E4438C Guide d'installation Page 101

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 362
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 100
Troubleshooting
Troubleshooting Analog Bus ADC Calibration Failures
1-89
Procedure for Signal Generators with Option UNJ
Run Self-Test 1100.
If Self-Test 1100 passes:
a. Turn the power off to the signal generator and remove the A15 Sampler (Option UNJ or Option
506), A16 Frac-N (Option UNJ or Option 506), and A18 Reference.
b. Using service software, run the Analog Bus ADC Calibration.
If the calibration passes:
a. Turn the power off to the signal generator.
b. Re-install the A15 Sampler (Option UNJ or Option 506).
c. Re-run the Analog Bus ADC Calibration.
If the calibration passes:
a. Re-install the A16 Frac-N (Option UNJ or Option 506).
b. Re-run the calibration.
If the calibration passes, replace the A18 Reference.
If the calibration fails, replace the A16 Frac-N (Option UNJ or Option 506).
If the calibration fails, replace the A15 Sampler (Option UNJ or Option 506).
If the calibration fails:
a. Turn the power off to the signal generator.
b. Re-install the A15 Sampler (Option UNJ or Option 506), A16 Frac-N (Option UNJ or Option
506), and A18 Reference.
c. Remove the A13 Output.
d. Re-run the Analog Bus ADC Calibration.
If the calibration passes, replace the A13 Output.
If the calibration fails, replace the A12 CPU.
If Self-Test 1100 fails, refer to “Self-Tests 11xx for the A12 CPU” on page 1-52.
Vue de la page 100
1 2 ... 96 97 98 99 100 101 102 103 104 105 106 ... 361 362

Commentaires sur ces manuels

Pas de commentaire