Agilent Technologies E4438C Guide d'installation Page 110

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 362
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 109
Troubleshooting
Troubleshooting Performance Test Problems
1-98
Pulse Modulation On/Off Ratio •Perform:
No adjustments
If failing, troubleshoot the following assemblies:
A18 Reference (fails at all frequencies)
A13 Output (fails only 4 GHz)
A14 Extended Frequency Output (Option 506)
(fails only > 4 GHz)
Burst Modulation On/Off Ratio Perform:
All Burst adjustments
If failing, troubleshoot the following assemblies:
A18 Reference (fails at all frequencies)
A13 Output (fails only 4 GHz)
A14 Extended Frequency Output (Option 506)
(fails only > 4 GHz)
Table 1-40 Troubleshooting Performance Test Problems
Performance test that is failing... Action to perform...
Vue de la page 109
1 2 ... 105 106 107 108 109 110 111 112 113 114 115 ... 361 362

Commentaires sur ces manuels