Assembly Replacement
A6 Power Supply
2-20
A6 Power Supply
Tools Required
T-10 driver
Removal Procedure
Refer to Figure 2-10 for this procedure.
1. Disconnect the power cord.
2. Remove the outer cover. Refer to “Outer and Inner Instrument Covers” on page 2-4.
3. Using the T-10 driver, remove the three flathead screws (1) and the ten pan-head screws (2)
from the A6 Power Supply’s sheet metal cover (3).
4. Disconnect the A24 Line Module cable (A24W1) from the A6 Power Supply (5).
5. Using the T-10 driver, remove the ten screws (4) that hold the A6 Power Supply (5)
in place.
6. Remove the A6 Power Supply.
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Torque all T-10 screws to 9 in-lbs.
3. Perform the post-repair adjustments and performance tests that pertain to this removal procedure.