Assembly Replacement
A15 Sampler (Option UNJ or Option 506)
2-36
A15 Sampler (Option UNJ or Option 506)
Tools Required
•none
Removal Procedure
Refer to Figure 2-18 for this procedure and to the connector/cable diagram located on the card cage.
1. Disconnect the power cord.
2. Remove the outer and inner-side covers. Refer to “Outer and Inner Instrument Covers” on page 2-4.
3. Locate Side-SLOT 3, lift the retention levers, and disconnect the A15 Sampler (Option UNJ or Option 506)
from the A19 Daughterboard.
4. Slide out the A15 Sampler (Option UNJ or Option 506).
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments and performance tests that pertain to this removal procedure.