Agilent Technologies PSG Manuel d'instructions Page 182

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 362
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 181
Command Reference
Communication Subsystem (:SYSTem:COMMunicate)
Chapter 4168
:SERial:RESet
Supported All
:SYSTem:COMMunicate:SERial:RESet
This event command resets the RS-232 buffer and will discard any unprocessed SCPI
input received by the RS-232 port.
*RST N/A
Range N/A
Key Entry
Reset RS-232
Remarks N/A
:SERial:TOUT
Supported All
:SYSTem:COMMunicate:SERial:TOUT <val>
:SYSTem:COMMunicate:SERial:TOUT?
This command sets the value for the RS-232 serial port time-out. If further input is not
received within the assigned time-out period while a SCPI command is being processed,
the command is aborted and the input buffer is cleared.
The variable <val> is entered in units of seconds.
*RST N/A
Range 10–60
Key Entry
RS-232 Timeout
Remarks The setting enabled by this command is not affected by signal
generator power-on, preset, or *RST.
Vue de la page 181
1 2 ... 177 178 179 180 181 182 183 184 185 186 187 ... 361 362

Commentaires sur ces manuels

Pas de commentaire