Agilent Technologies PSG Manuel d'instructions Page 257

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 362
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 256
Command Reference
Frequency Subsystem ([:SOURce])
Chapter 4 243
:ROSCillator:SOURce:AUTO
Supported All except signal generators with Option UNJ
[:SOURce]:ROSCillator:SOURce:AUTO ON|OFF|1|0
[:SOURce]:ROSCillator:SOURce:AUTO?
This command enables or disables the ability of the signal generator to automatically
select between the internal and an external reference oscillator.
ON (1) This choice enables the signal generator to detect when a valid
reference signal is present at the 10 MHz IN connector and
automatically switches from internal to external frequency reference.
OFF (0) This choice selects the internal reference oscillator and disables the
switching capability between the internal and an external frequency
reference.
*RST 1
Key Entry
Ref Oscillator Source Auto Off On
Remarks N/A
Choices ON OFF 1 0
Vue de la page 256