Agilent Technologies PSG Manuel d'instructions Page 207

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 362
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 206
Command Reference
Status Subsystem (:STATus)
Chapter 4 193
:QUEStionable:CALibration:CONDition
Supported PSG-A Series
:STATus:QUEStionable:CALibration:CONDition?
This command returns the decimal sum of the bits in the Data Questionable Calibration
Condition Register. For example, if the DCFM or DC
Φ
M zero calibration fails (bit 0), a
value of 1 is returned.
*RST N/A
Range 0–32767
Key Entry N/A
Remarks Refer to Data Questionable Calibration Status Group” on page 133
and “Data Questionable Calibration Condition Register” on page 134
for more information.
The data in this register is continuously updated and reflects the
current conditions.
:QUEStionable:CALibration:ENABle
Supported PSG-A Series
:STATus:QUEStionable:CALibration:ENABle <value>
:STATus:QUEStionable:CALibration:ENABle?
This command determines what bits in the Data Questionable Calibration Event
Register will set the calibration summary bit (bit 8) in the Data Questionable Condition
Register.
The variable <value> is the sum of the decimal values of the bits that you want to
enable.
*RST N/A
Range 0–32767
Key Entry N/A
Remarks Refer to Data Questionable Calibration Status Group” on page 133
and “Data Questionable Calibration Event Enable Register” on
page 135 for more information.
Vue de la page 206
1 2 ... 202 203 204 205 206 207 208 209 210 211 212 ... 361 362

Commentaires sur ces manuels

Pas de commentaire