Command Reference
Frequency Subsystem ([:SOURce])
Chapter 4 241
:FREQuency[:CW]
Supported All
[:SOURce]:FREQuency[:CW] <val><unit>
[:SOURce]:FREQuency[:CW]?
This command sets the signal generator’s output frequency for the CW and FIXed
frequency modes.
Key Entry
Frequency
Remarks Refer to “:FREQuency:MODE” on page 237 for setting the frequency
type.
:PHASe:REFerence
Supported All
[:SOURce]:PHASe:REFerence
This command sets the current output phase as a zero reference.
*RST N/A
Range N/A
Key Entry
Phase Ref Set
Remarks Subsequent phase adjustments are set relative to the new reference.
*RST 20 GHz Models:
+
2.0000000000000E
+
10
40 GHz Models:
+
4.0000000000000E
+
10
Range 20 GHz Models: 100kHZ–20GHZ 40 GHz Models:100kHZ40GHZ