Sink Eye Diagram Tests 18
Agilent U7232A DisplayPort Electrical Performance Compliance Test Application 161
10 The following select file dialog box appears. Select your coefficient file
(*.equ) and click Open. The test will run based on the your user
defined coefficient file.
PASS Condition
The following table and figure define the mask for the eye measurements.
There can be no signal trajectories entering into the mask. Table 21 shows
the voltage and time coordinates for the mask used for the eye diagram.
Figure 52 Selecting User Defined File