Agilent Technologies DisplayPort Manuel de service Page 216

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 246
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 215
216 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
24 Calibrating the Infiniium Oscilloscope and Probe
3 Referring to Figure 81 below, perform the following steps:
a Click on the Utilities>Calibration menu to open the Calibration
window.
4 Referring to Figure 82 below, perform the following steps to start the
calibration:
a Uncheck the Cal Memory Protect checkbox.
b Click the Start button to begin the calibration.
Figure 81 Accessing the Calibration Menu.
Vue de la page 215
1 2 ... 211 212 213 214 215 216 217 218 219 220 221 ... 245 246

Commentaires sur ces manuels

Pas de commentaire