Agilent Technologies 11867A? Manuel de service Page 133

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 214
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 132
Chapter 8 135
Concepts
Stimulus Response Measurement Concepts
Concepts
level are dB, indicating that this is a relative measurement.
To make a reflection measurement accurately, it is necessary to perform
an open/short/load calibration. An open, short, and load are included in
the Stimulus Response Calibration Kit, Option SRK.
Press
Mode, Stimulus/Response, Return Loss. Set the desired start and
stop frequencies. Press
Calibrate and follow the instructions.
After the calibration is complete, connect the DUT to the RF OUTPUT
connector to make your return loss measurement. The marker readout
returns the amplitude values in both return loss and VSWR.
Vue de la page 132
1 2 ... 128 129 130 131 132 133 134 135 136 137 138 ... 213 214

Commentaires sur ces manuels

Pas de commentaire