9-4
Clock Recovery Commands
CLBandwidth
CLBandwidth
Command
(83496A Option
300)
:CRECovery{1 | 3}:CLBandwidth <bandwidth>
This 83496A Option 300 command sets or queries an 83496A Option 300 module’s loop band-
width. You must issue the LBWMode FIXed command before using the CLBandwidth com-
mand. A settings conflict error is reported if the module’s loop bandwidth mode is set to be
rate dependent (RDEPendent). Refer to “LBWMode” on page 9-6. The loop bandwidth can be
any bandwidth within 30 kHz to 10 MHz specified to 3 significant digits. The default setting is
60 kHz.
Restrictions Software revision A.04.20 and above.
Example 10 OUTPUT 707; ":CRECOVERY1:CLBANDWIDTH 1.7E6"
Query :CRECovery{1 | 3}:CLBandwidth?
Returned Format [:CRECovery{1 | 3}:CLBandwidth] <bandwidth><NL>
Table 9-2. Module Data Rates
Rate
(Mb/s)
83491
83492
83493
83494
83494
Option 103
83494
Option 106
83494
Option 107
83495
83496
83496
Option 200
Trigger on data
••••
155.52
••••
622.08
••••
1062.50
••
1250.00
•••
2125.00
••
2488.32
•••
2500.00
•••
2666.06
•• ••
9953.28
••
10312.50
••
10664.23
••
10709,225
••
9.953 Gb/s–
11.32 Gb/s
••
Continuous
••