Agilent Technologies 86100A Manuel de service Page 336

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 364
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 335
24-6
Trigger Commands
SLOPe
SLOPe
Command :TRIGger:SLOPe {POSitive | NEGative}
This command specifies the slope of the edge on which to trigger.
Query :TRIGger:SLOPe?
The query returns the slope for the trigger.
Returned Format [:TRIGger:SLOPe] {POSitive | NEGative}<NL>
Example 10 OUTPUT 707; ":TRIGger:SLOPe POSitive"
SOURce
Command :TRIGger:SOURce {FPANel | FRUN | LMODule | RMODule}
This command selects the trigger input. Front Panel, Left Module, and Right Module are
inputs from the front panel of the instrument. Free Run is internally generated, and is not
affected by the settings of gates, level, slope, bandwidth, or hysteresis.
<trigger> Front PANel, Left MODule, and Right MODule are inputs on the front of the instrument. Fre-
eRUN is internally generated and is unaffected by the settings for gated, level, slope, band-
width or hysteresis.
Query :TRIGger:SOURce?
The query returns the current trigger source of the current mode.
Returned Format [:TRIGger:SOURce] <trigger><NL>
Vue de la page 335
1 2 ... 331 332 333 334 335 336 337 338 339 340 341 ... 363 364

Commentaires sur ces manuels

Pas de commentaire