To create a branching sequence
Each sequence step can be branched to different steps depending on the
result of the mask testing.
1 Press Print/Utility twice.
2 Press the Sequence Admin softkey.
3 Press the Define Sequence softkey.
4 Press the Step softkey, or turn the knob closest to the Cursors key,
to select the step number to be modified.
5 Press the Define Step softkey.
6 Press the Branching softkey.
A softkey menu with five softkey choices appears. Four of them are used to
define branching.
Pass, Fail Min, Fail Max Selects the branching destination for
different test results: pass, fail minimum limit, or fail maximum limit. The
destination step is selected by turning the knob closest to the Cursors key.
Possible selections are step number, Next, or Same.
Default Branching Sets the branching destination for all test results
to Next.
Previous Menu Returns you to the previous softkey menu.
7 Press the branching destination softkey corresponding to the test
result.
8 Turn the knob closest to the Cursors key to change the branching
destination.
Operating the Test Automation Module
To create a branching sequence
47