Creating a DC voltage testing mask
A mask can be used to test the proper voltage of a DC signal. The shape of
the mask is very simple: just two horizontal straight lines. By varying the
Volts/div knob and the position setting, the same DC voltage testing mask can
be used to test any DC signals. To create the DC testing mask, you must
have the testing limits. The testing limits are the upper voltage limit and the
lower voltage limit. To maximize the accuracy, the two limit lines should
cover as much vertical resolution as possible. Due to the calibrated vertical
verniers and vertical position of the HP 54600-series oscilloscopes, the
Volts/div can be expanded in small increments and remain calibrated. An
example mask is shown below.
Example mask for testing the voltage of a DC signal
Figure 13
Solutions
Creating a DC voltage testing mask
65