Storing multiple sequences for multiple devices
The Test Automation Module can store 100 sequence steps and 40 mask
templates. In most cases, a test sequence uses less than 100 steps and
40 mask templates. Therefore, you can store multiple sequences in one Test
Automation Module for testing multiple devices.
The guideline for storing multiple sequences is that each sequence resides in
its own section. An extra header step is added to each sequence. The
header step should contain the following information:
A label that clearly identifies the corresponding test sequence for the
header step.
The pass branching which should branch to the first step of the
corresponding test sequence when the header step is selected.
These header steps should be placed together at the beginning of the
sequencing memory, from step 1, like a table of contents. When the operator
wants to start one of the sequences in the module, the following procedure
should be performed.
1 Press Setup .
2 Press the Sequence Menu softkey.
3 Press the Reset softkey to reset the step number to 1.
4 Press the Step softkey, or turn the knob closest to the Cursors key,
until the header step of the desired sequence appears.
Since all of the first steps of the sequences are gathered around the top of
the module’s memory, you can easily scroll through them and find the one
that you want to execute.
The figure 24 shows how to place the sequences in the module’s memory.