Measure the Detector Constant
When "Measure the Detector Constant" is selected, the HP 3048A system
automatically measures the Phase Detector Constant. Measuring the
Phase Detector Constant is the most accurate method for calibrating the
detector. The system calibrates the Phase Detector Constant by generating
a beatnote frequency at approximately 10% of the entered Peak Tuning
Range. The maximum Peak Tuning Range the system will except is
500 kHz (or 200 MHz if a supported RF Analyzer is configured in the
system).
If the beatnote is less than 1 kHz, the system measures the slope of the
zero crossings to determine the Phase Detector Constant. If the beatnote is
greater than or equal to 1 kHz, the system measures the amplitude of the
fundamental of the beatnote signal and its odd harmonics to determine
the Detector Constant.
Detector Constant^
kHz
= Slope
Bea
tnote
Detector Constant^ ^^z = H\
3H%
+ 5#5
1H-;
■ ■
where H\ = Peak amplitude of fundamental of beatnote
and H
n
= Amplitude of harmonics of beatnote
Considerations for the Phase Detector Constant
The HP 11848A External Noise Input must be attenuated as needed to
limit the Phase Detector Constant to less than 1 V/Rad to keep from
overloading the HP 11848A. The beatnote must be DC coupled to the
input for the Phase Lock Loop to operate.
Calibr Process: $N w/PLL 2-47