Store File: stores a Test File onto the specified Mass Storage Media. The
Test File contains the defined measurement parameters and the Calibration
Constants derived during the last measurement.
Considerations for Test Files
The System Configuration Table is not stored with the Test File. When
loading a Test File from the Mass Storage Media, it is possible for
the Source Control Diagram to be configured incorrectly. If the Source
Control Diagram contained in a Test File that is being loaded specifies an
instrument that is not listed in the current System Configuration Table, the
HP 3048A will not be able to include the specified instrument in the Source
Control Diagram. Instead, the HP 3048A will specify the instrument that
is currently listed in the System Configuration Table, or if an appropriate
instrument is not listed, the Users Source specification will be selected.
Copying the Test Files using the copy routine provided in the Basic
Operating System will not update the Test File Directory and may cause
loss of the data. Copy Test Files from one disc to another by loading the
Test File into the System, then storing the Test File on to the other disc.
The Parm_number shown for each Test File represents the media file name
and the order in which the Test Files were stored.
2-124 Test Files
reu.05SEP89