Agilent Technologies HP 8642A/B Spécifications Page 168

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 453
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 167
Considerations for the Derive From Single-Sided
Spur Method
The power level at the AM detector input must remain the same for
both calibration measurements and the noise measurements. For optimum
results, a power level of +10 dBm is recommended when using the
HP 33330C AM Detector.
Requires an RF Spectrum Analyzer capable of measuring the calibration
spur at the desired carrier frequency. Requires a second RF Generator that
can be set to an offset frequency of from 1 Hz to 100 kHz (or 20 mHz if
an RF Analyzer is configured in the system).
Calibration is performed under actual measurement setup conditions.
The non-linearity and harmonics of the AM detector are calibrated out.
2-100 Calibr Process: AM Noise
Vue de la page 167
1 2 ... 163 164 165 166 167 168 169 170 171 172 173 ... 452 453

Commentaires sur ces manuels

Pas de commentaire