4 Triggering
120 InfiniiVision Oscilloscopes User’s Guide
If don't care (0xXX) is selected for data, the data will be ignored. The trigger will
always occur on the 17th clock for 7-bit addressing or 26th clock for 10-bit
addressing.
c If you have selected a three-byte trigger, press the Data2 value softkey and turn the
Entry knob to select the 8-bit data pattern on which to trigger.
NOTE
To display I
2
C serial decode, see page 310.